Technical specification for oxygen analyzer
|
tarix | 27.02.2023 | ölçüsü | 12,26 Kb. | | #85741 |
| 05-10-Oxygen Analyzer
TECHNICAL SPECIFICATION FOR OXYGEN ANALYZER
-
1 unit of oxygen analyzer will be purchased
-
Measurement range must fulfill 0-100 %
-
Accuracy must be minimum 0.01 % O2
-
Technology should be paramagnetic sensing or equivalent
-
A flowmeter 500 ml/min for sample inlet should be given
-
Configurable alarms must accompany the system
-
Mounting Rack Mount must be available
-
8 Relay Contacts w/connectors should be available
-
Sample inlet must be 1/8" NPT male
-
Sample outlet should be 1/4" NPT female
-
Zero drift per week should be maximum <0.01 % O2
-
The system must be working with sample gas temperature 5-40 °C
-
It should be suitable for 220 V 50/60 Hz electrical current